Title: Automatic radioxenon analyzer for CTBT monitoring

Technical Report ·
DOI:https://doi.org/10.2172/431123· OSTI ID:431123

Over the past 3 years, with support from US DOE`s NN-20 Comprehensive Test Ban Treaty (CTBT) R&D program, PNNL has developed and demonstrated a fully automatic analyzer for collecting and measuring the four Xe radionuclides, {sup 131m}Xe(11.9 d), {sup 133m}Xe(2.19 d), {sup 133}Xe (5.24 d), and {sup 135}Xe(9.10 h), in the atmosphere. These radionuclides are important signatures in monitoring for compliance to a CTBT. Activity ratios permit discriminating radioxenon from nuclear detonation and that from nuclear reactor operations, nuclear fuel reprocessing, or medical isotope production and usage. In the analyzer, Xe is continuously and automatically separated from the atmosphere at flow rates of about 7 m{sup 3}/h on sorption bed. Aliquots collected for 6-12 h are automatically analyzed by electron-photon coincidence spectrometry to produce sensitivities in the range of 20-100 {mu}Bq/m{sup 3} of air, about 100-fold better than with reported laboratory-based procedures for short time collection intervals. Spectral data are automatically analyzed and the calculated radioxenon concentrations and raw gamma- ray spectra automatically transmitted to data centers.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
Sponsoring Organization:
USDOE, Washington, DC (United States)
DOE Contract Number:
AC06-76RL01830
OSTI ID:
431123
Report Number(s):
PNNL-11424; ON: DE97050869; TRN: 97:003206
Resource Relation:
Other Information: PBD: Dec 1996
Country of Publication:
United States
Language:
English