Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

THE DISTRIBUTION OF IONIZATION ALONG $gamma$-RAY BEAM AND REPRODUCING THE ROENTGEN WITH NORMAL IONIZATION CHAMBERS (in Russian)

Journal Article · · Atomnaya Energ.
OSTI ID:4261833
True ionization from gamma radiation and resulting x radiation may be found from standard ionization chamber measurements by introducing a displacement factor, ~0.35 (R/sub max/P), for distance from the measuring electrode center. The displacement magnitude was determined and used in studies of Cs/sup 137/ and Co/up 60/ gamma radiation. The corrections for the displacement for Cs/sup 137/ and Co/sup 60/ were 0.996 and 0.992, respectively. (R.V.J.)
Research Organization:
Originating Research Org. not identified
NSA Number:
NSA-13-018051
OSTI ID:
4261833
Journal Information:
Atomnaya Energ., Journal Name: Atomnaya Energ. Vol. Vol: 6
Country of Publication:
Country unknown/Code not available
Language:
Russian

Similar Records

Cylindrical ionization 4. pi. -chamber for testing the nuclide activity of reference gamma sources
Journal Article · Fri Oct 31 23:00:00 EST 1975 · Meas. Tech. (USSR) (Engl. Transl.); (United States) · OSTI ID:7156437

AN APPROXIMATE CALCULATION OF THE MEAN ENERGY OF THE ELECTRONS EJECTED BY $gamma$-RAYS IN AN IONIZATION CHAMBER
Journal Article · Mon Aug 01 00:00:00 EDT 1960 · Atomnaya Energ. · OSTI ID:4111652

ON THE MEASUREMENT OF THE DOSE OF GAMMA AND BETA RADIATION WITH THE AID OF THIMBLE IONIZATION CHAMBERS
Journal Article · Sun May 01 00:00:00 EDT 1960 · Izmeritel'naya Tekh. · OSTI ID:4166250