A SCINTILLATION TRANSISTORIZED ALPHA A-C OPERATED "POPPY" MONITOR
Technical Report
·
OSTI ID:4254379
A 17-pound, semi-portable, scintillation, transistorized alpha monitor was designed and fabricated, The instrument has both a loud-speaker and a count- ratemeter output indication. The count-rate-meter ranges are 0-1000, 0-10,000, and 0-100,000 C/M with an average probe geometry for Pu/sup 239/ alpha particles of about 18 per cent. The background count-rate is stable at less than three C/M for weeks at a time with no adjustments necessary. All transistor circuits are of the plug-in, interchangcable, printed circuit type. The loudspeaker volume output is sufficient for most plant locations, This instrument was designed primarily for alpha detcction but it may be converted to beta-gamma or neutron detection by proper crystal detector selection. (auth)
- Research Organization:
- General Electric Co. Hanford Atomic Products Operation, Richland, Wash.
- DOE Contract Number:
- W-31-109-ENG-52
- NSA Number:
- NSA-13-013446
- OSTI ID:
- 4254379
- Report Number(s):
- HW-57506
- Country of Publication:
- United States
- Language:
- English
Similar Records
A SENSITIVE GAMMA SCINTILLATION TRANSISTORIZED MONITOR
A SLOW AND FAST NEUTRON SCINTILLATION COUNT-RATE AND DOSE-RATE METER
PORTABLE RADIATION INSTRUMENTATION STANDARDIZATION
Technical Report
·
Mon Apr 20 23:00:00 EST 1959
·
OSTI ID:4246700
A SLOW AND FAST NEUTRON SCINTILLATION COUNT-RATE AND DOSE-RATE METER
Technical Report
·
Fri Aug 01 00:00:00 EDT 1958
·
OSTI ID:4327804
PORTABLE RADIATION INSTRUMENTATION STANDARDIZATION
Technical Report
·
Fri Jul 01 00:00:00 EDT 1960
·
OSTI ID:4116191