A SCINTILLATION TRANSISTORIZED ALPHA-BETA-GAMMA HAND AND SHOE COUNTER

Technical Report ·
OSTI ID:4216320
The instrument was developed to provide an efficient, sensitive, reliable method of determining the extent of radioactive contamination on the bands, shoes, and clothing of personnel employed at the Hanford Atomic Products Operation, General Electric Company, Richland, Washington. The instrument can simultaneously detect and indicate radioactive contamination levels for alpha and beta particles and gamma photons which are produced by disintegration of radioisotopes. The instrument replaced, in one unit, two separate instruments of vacuum tube types designed for the same purpose. With the new instrument, complete personnel contamination determination is accomplished in less than one- half the time previously used. The new instrument has more than twice the detection sensitivity of either of the previous vacuum tube instruments. Multiplier phototubes and scintillating crystal detectors are used in the new instrument to achieve a reliability not obtainable with the air proportional chambers and Geiger-Mueller tubes used for alpha and beta-gamma detecting, respectively, in the two older instruments. (auth)
Research Organization:
General Electric Co. Hanford Atomic Products Operation, Richland, Wash.
DOE Contract Number:
AT(45-1)-1350
NSA Number:
NSA-14-005330
OSTI ID:
4216320
Report Number(s):
HW-60631
Country of Publication:
United States
Language:
English

Similar Records

AN ALPHA, BETA, GAMMA HAND AND SHOE COUNTER-MODEL II
Technical Report · Sun Jun 01 00:00:00 EDT 1958 · OSTI ID:4314550

A METHOD FOR BETA-GAMMA HAND AND SHOE COUNTING UNDER VARYING BACKGROUND CONDITIONS
Technical Report · Tue Oct 01 00:00:00 EDT 1963 · OSTI ID:4080398

PERSONNEL RADIOACTIVITY MONITOR
Journal Article · Fri May 01 00:00:00 EDT 1959 · Strumentazione e automazione · OSTI ID:4235368