Complexometric titration of rare earth elements. Dissolution of the rare earth oxalate with ethylenediaminetetraacetic acid and back titration with magnesium sulfate
The processes for dissolution of the oxalate and complexometry to determine the small amounts of rare earth rapidly and accurately, were studied. Crystalline rare earth oxalate was precipitated and filtered. The precipitate was dissolved completely with EDTA in alkaline solution and the excess of EDTA in solution was back titrated with standard magnesium sulfate solution. By this procedure, rare earths in the range of a content about 1 to 5 mg, could be determined rapidly with a good accuracy.
- Research Organization:
- Kyushu Univ., Fukuoka
- NSA Number:
- NSA-14-003519
- OSTI ID:
- 4200576
- Journal Information:
- Bull. Chem. Soc. Japan, Journal Name: Bull. Chem. Soc. Japan Vol. Vol: 32
- Country of Publication:
- Country unknown/Code not available
- Language:
- English
Similar Records
COMPLEXOMETRIC TITRATION OF YTTRIUM AND THE LANTHANONS. II. METHODS FOR THEIR DETERMINATION IN OXALATES
Dissolution and separation of rare earth metals
THE ANALYTICAL METHODS FOR CALCIUM-SILICON CONTAINING A LARGE AMOUNT OF RARE EARTHS
Journal Article
·
Thu Oct 31 23:00:00 EST 1963
· Talanta (England)
·
OSTI ID:4147127
Dissolution and separation of rare earth metals
Patent
·
Tue May 12 00:00:00 EDT 2020
·
OSTI ID:1650890
THE ANALYTICAL METHODS FOR CALCIUM-SILICON CONTAINING A LARGE AMOUNT OF RARE EARTHS
Journal Article
·
Wed Oct 31 23:00:00 EST 1962
· Bunseki Kagaku (Japan)
·
OSTI ID:4156118