Title: Application of a low energy x-ray spectrometer to analyses of suspended air particulate matter

Conference ·
OSTI ID:4192178

A semiconductor detector x-ray spectrometer has been constructed for the analysis of elements in air particulate specimens. The excitation radiation is provided, either directly or indirectly, using a low power (40 watts) Ag anode x- ray tube. Less than 100 ng for most of the elements in the range Mg $Yields$ Zr, Pb are easily detected within two 1-minute counting intervals. A calibration technique for light element analysis and an experimental method which compensates for particle size effects are discussed. (auth)

Research Organization:
California Univ., Berkeley (USA). Lawrence Berkeley Lab.
DOE Contract Number:
W-7405-ENG-48
NSA Number:
NSA-33-000778
OSTI ID:
4192178
Report Number(s):
LBL-4402; CONF-750831-5
Resource Relation:
Conference: 24. annual conference on applications of x-ray analysis, Denver, Colorado, USA, 6 Aug 1975; Other Information: Orig. Receipt Date: 30-JUN-76