ION-CHAMBER RESPONSE IN HIGH-LEVEL RADIATION FIELD
Journal Article
·
· Nucleonics (U.S.) Ceased publication
OSTI ID:4190290
An adaptation of an ORNL ion chamber is described. The instrument was modified to meet the criteria set by an Air Force committee for the standardlzation of radiation measurement techniques. These criteria for satisfactory instrument performance are: (1) as nondirectional as possible; (2) known temperature response between 0 and 100 deg C; (3) known neutron response as low as possible; (4) capable of withstanding 10/sup 12/ ergs/g (C); (5) sturdy and simple construction; (6) linear to plus or minus 1/2% at 10/sup 5/to 10/sup 10/ ergs/g (C)/hr. All these were met but the last. The results indicate that the chamber can be used in dose rates up to 4 x 10/x sup 10/ ergs/g (C)/hr with a 5 kv collecting potential at 1,000 mm absolute pressure and a 2 x 10/sup 10/ ergs/ g (C)/hr with 3 kv at 760 mm. (B.O.G.)
- Research Organization:
- General Dynamics Corp., Fort Worth, Tex.
- NSA Number:
- NSA-14-013888
- OSTI ID:
- 4190290
- Journal Information:
- Nucleonics (U.S.) Ceased publication, Journal Name: Nucleonics (U.S.) Ceased publication Vol. Vol: 18, No. 5; ISSN NUCLA
- Country of Publication:
- Country unknown/Code not available
- Language:
- English
Similar Records
MODEL IV-GRAPHITE, CARBON DIOXIDE 4-CC IONIZATION CHAMBER
MO-FG-CAMPUS-IeP1-05: New Ionization Chamber Dosimetry of Absorbed Dose to Water in Diagnostic KV X-Ray Beams
IN-PILE IONIZATION CHAMBERS
Technical Report
·
Mon Feb 29 23:00:00 EST 1960
·
OSTI ID:4166345
MO-FG-CAMPUS-IeP1-05: New Ionization Chamber Dosimetry of Absorbed Dose to Water in Diagnostic KV X-Ray Beams
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22653887
IN-PILE IONIZATION CHAMBERS
Technical Report
·
Sun Oct 30 23:00:00 EST 1960
·
OSTI ID:4216485