GONADAL EXPOSURE DOSE TO ADULTS IN DIAGNOSTIC RADIOGRAPHY
Technical Report
·
OSTI ID:4180332
Measurements of gonadal exposure dose to adults in diagnostic radiography were made under a wide variety of conditions, using a phantom. The measurements of both skin and gonadal dose are presented in 46 tables, for both males and females, for nineteen of the views most commonly taken in medical radiography. The tables include data showing the variation of exposure dose with such quantities as target-film distance, filtration, peak kilovoltage, and field size. (auth)
- Research Organization:
- National Research Council of Canada. Div. of Applied Physics, Ottawa
- NSA Number:
- NSA-14-021278
- OSTI ID:
- 4180332
- Report Number(s):
- NRC-5377; APXNR-809
- Country of Publication:
- Country unknown/Code not available
- Language:
- English