CALORIMETRIC DETERMINATION OF ABSORBED DOSE AND GFe +++ OF THE FRICKE DOSIMETER WITH 10-Mev AND 20-Mev ELECTRONS
The construction and operation details of a local absorbed dose microcalorimeter are presented. This calorimeter measures the absorbed dose rate at any desired point within an irradiated medium. The measurement is specific to a sufficiently limited region (2.0 cm in dia, 0.3 cm deep) that the measurement is essentially at a point and no integration or averaging process is necessary to qualify either the absorbed dose, or its rate. The system is sufficiently sensitive to provide about 1% accuracy at dose rates of 50 rads/min. The application of this calorimeter to 10-Mev and 20-Mev incident electrons is described. The use of a secondary standard (Victoreen 25-r chamber) under specified conditions is given together with details of calibration for use with high-energy electrons. The influence of the polarization effect was noted. The Fricke ferrous sulfate dosirieter was also calibrated for these energies of electrons. Samples of dosimeter solution were exposed in a geometry identical with that of the small sensitive region of the calorimeter. A value of G/sub Fe// sup 3+/ of 15.32 plus or minus 0.34 was obtained for 10-Mev electrons, and a value of 15.17 plus or minus 0.28 for 20-Mev electrons. (auth)
- Research Organization:
- Sloan-Kettering Inst., New York
- Sponsoring Organization:
- USDOE
- NSA Number:
- NSA-18-002072
- OSTI ID:
- 4156518
- Report Number(s):
- TID-19171
- Journal Information:
- Radiation Res., Journal Name: Radiation Res. Vol. Vol: 20
- Country of Publication:
- Country unknown/Code not available
- Language:
- English
Similar Records
A calorimetric determination of the oxidation yield of the Fricke dosimeter at high dose rates of electrons
Absorbed gamma radiation dose in the Fricke dosimeter
The Relative Biological Efficiency of 20-Mev Electrons and 180-Kvp X-Rays in Escherichia coli Inactivation
Journal Article
·
Sun Dec 31 23:00:00 EST 1961
· Journal of Physical Chemistry
·
OSTI ID:4818794
Absorbed gamma radiation dose in the Fricke dosimeter
Journal Article
·
Wed Dec 31 23:00:00 EST 1980
· Annu. Book ASTM Stand.; (United States)
·
OSTI ID:5549850
The Relative Biological Efficiency of 20-Mev Electrons and 180-Kvp X-Rays in Escherichia coli Inactivation
Journal Article
·
Thu Sep 01 00:00:00 EDT 1960
· Radiation Research
·
OSTI ID:4146688