In order to establish the feasibility of using two colorimetric met;iods for the determination of uranium interchangeably, according to the interferences encountered in a particular sampte. results were obtained by each of the methods and compared. The dibenzoyl methane method and the ethyl acetate-ammonium thiocyanate procedure were compared on the basis of values secured on the same day. on different days, on an analysis of the variance. and on an analysis of the residual error for the methods on different days. On the basis of the findings of these tests. it is concluded that the two methods can be used interchangeably to determine the uranium content of the ethyl acetate extracts of samples. Since the interferences in the two methods are different. the uranium content of a variety of materials can be determined without additional separations being required. (auth)