The use of threshold detector foils to measure the external neutron spectrum of the Livermore Pool Type Reactor and the KEWB Reactor is described. The foils employed are Pu/sup 239/ to measure the total fast flux, U/sup 238/for the region above l.5 Mev, S/sup 32/ above 2.5 Mev, and Au/sup 197/ for thermal neutrons. Scintillation counters are used to determine the activity induced in the detectors. The foil-counting system is calibrated by using a Cockcroft- Walton accelerator as a fast neutron source. The results indicate a fission type spectrum at each reactor. (auth)