Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

POLAROGRAPHIC DETERMINATION OF HEXAVALENT URANIUM IN URANIUM TETRAFLUORIDE (in Japanese)

Journal Article · · J. Atomic Energy Soc. Japan
DOI:https://doi.org/10.3327/jaesj.2.313· OSTI ID:4141544
An analytical method is proposed for the polarographic determination of uranium(VI) in uranium tetrafluoride. A sample weighing about 500 mg is treated with 20 ml of 0.1M NaF solution using a magnetic stirrer, and is subjected to centrifugation. Two milliliters of the supernatant solution and 5 ml of supporting electrolyte solution (1.3 M HOAc + 0.7 M NH/sub 4/OAc + 0.2 M BeSO/sub 4/) are taken into a 10-ml volumetric flask and the resulting solution is diluted to the mark with water. Uranium(VI) in the solution is determined polarographically by referring to the standard calibration curve. For the purpose of suppressing the dissolution of uranium(IV) into the solution, NaF is used at the time of the sample digestion. The acetate buffer solution keeps the pH of the solution at 5 and Be/sup 2+/ ion removes the interference of F/sup -/ ion in the polarographic measurements. Al/sup 3+/ can be used instead of Be/sup 2+/ for the same purpose. By this method as small as 0.05% of uranium(VI) in uranium tetrafluoride can be readily determined. (auth)
Research Organization:
Japan Atomic Energy Research Inst., Tokyo
Sponsoring Organization:
USDOE
NSA Number:
NSA-14-018868
OSTI ID:
4141544
Journal Information:
J. Atomic Energy Soc. Japan, Journal Name: J. Atomic Energy Soc. Japan Vol. Vol: 2
Country of Publication:
Country unknown/Code not available
Language:
Japanese