BRIGHT FIELD PHOTOGRAPHY IN A HYDROGEN BUBBLE CHAMBER
The performance of bright and dark field optics in a bubble chamber is analyzed. A modification of the usual bright field scheme is proposed, so that very high contrasts may be achieved even in liquid hydrogen. The new scheme can also be used in a single-window chamber. Results of tests of the scheme are presented to support this modification. Photography of bubbles smaller than the resolution limit is discussed; in this case a dark field scheme shows a clear advantage. (auth)