TRANSMISSION OPTICS OF FOCUSED ION BEAMS USED IN MASS SPECTROMETRY

Journal Article · · Rev. Sci. Instr.
DOI:https://doi.org/10.1063/1.1716859· OSTI ID:4100691
Current density profiles and angular spreading of low intensity positive ion beams are measured with scanning slits in an evacuated beam analyzer. Experimental measurements are made of spherical aberration and transverse thermal velocity at a crossover focused by an ion gun. Spherical aberration is determined by an extrapolation technique which apparently has not been used before. Beam constants were evaluated for an ion gun used with V type, flat ribbon, and multiple filaments. Each type of surface ionization filament produces a beam of elliptical cross section. Expressions are given for the variations of current density in a plane normal to and moved along the beam axis in field-free image space, and for the beam fraction transmitted through a rectangular aperture. Application of the results to estimate the degree of voltage and mass discriminations is discussed. (auth)
Research Organization:
Knolls Atomic Power Lab., Schenectady, N.Y.
NSA Number:
NSA-15-006182
OSTI ID:
4100691
Journal Information:
Rev. Sci. Instr., Journal Name: Rev. Sci. Instr. Vol. Vol: 31
Country of Publication:
United States
Language:
English

Similar Records

Aberration of a negative ion beam caused by space charge effect
Journal Article · Sun Feb 14 23:00:00 EST 2010 · Review of Scientific Instruments · OSTI ID:22053842

Optimizing performance of double-focusing mass spectrometers. 1. Determination and use of total aberrations
Journal Article · · Energy and Fuels; (United States) · OSTI ID:5048743

Use of the focusing multi-slit ion optical system at RUssian Diagnostic Injector (RUDI)
Journal Article · Tue Feb 14 23:00:00 EST 2012 · Review of Scientific Instruments · OSTI ID:22066311