COBALT-60 DEPTH-DOSE CORRECTIONS AS DETERMINED BY TRANSMISSION-DOSE MEASUREMENTS
A method for correcting the central-axis depth dose for tissue density variations is described. The only equipment needed is a Victoreen condenser r- meter and an inexpensive collimator of simple construction. The technic, which is based upon patient transmission, involves a transit dosimeter reading of radiation transmitted by the patient during the course of a routine treatment, a transit dosimeter reading made under identical conditions but with the patient removed, and the patient' s anteroposterior diameter. With the aid of data preserted above, which should be applicable to all cobalt teletherapy apparatus, the appropriate correction factors are readily obtained. (auth)
- Research Organization:
- Yeshiva Univ., New York; and Bronx Municipal Hospital Center, New York
- Sponsoring Organization:
- USDOE
- NSA Number:
- NSA-15-021004
- OSTI ID:
- 4088601
- Journal Information:
- Radiology (U.S.), Journal Name: Radiology (U.S.) Vol. Vol: 76; ISSN RADLA
- Country of Publication:
- Country unknown/Code not available
- Language:
- English
Similar Records
COBALT-60 DEPTH-DOSE CORRECTION AS DETERMINED BY TRANSMISSION DOSE MEASUREMENTS
Modification of a standard cobalt-60 unit for total body irradiation at 150 cm SSD
Nationwide survey of cobalt-60 teletherapy. Final report
Journal Article
·
Sat Dec 31 23:00:00 EST 1960
· Radiology (U.S.)
·
OSTI ID:4117052
Modification of a standard cobalt-60 unit for total body irradiation at 150 cm SSD
Journal Article
·
Fri Jun 01 00:00:00 EDT 1984
· Int. J. Radiat. Oncol., Biol. Phys.; (United States)
·
OSTI ID:6865393
Nationwide survey of cobalt-60 teletherapy. Final report
Technical Report
·
Fri Aug 01 00:00:00 EDT 1980
·
OSTI ID:6942939