SYNCHROCYCLOTRON 200-METER FLIGHT PATH NEUTRON VELOCITY SPECTROMETER

Journal Article · · Rev. Sci. Instr.
DOI:https://doi.org/10.1063/1.1718804· OSTI ID:4079466
The synchrocyclotron neutron time-of-flight spectrometer uses a 200-m flight path and two 2000-channel analyzers The modifications and improvements described have been made as a result of operating experience. The operation uses approximates 20 nsec bursts at an instantaneous intensity approximates 5 x 10/ sup 18/ evaporation neutrons per second and a 60-cps repetition rate. The resolution for energies above 1 kev is limited by the time analyzers, which have 100-nsec detection channel widths, giving a resolution of 0.5 nsec/m in the kev region. (auth)
Research Organization:
Columbia Univ., New York
Sponsoring Organization:
USDOE
NSA Number:
NSA-18-014275
OSTI ID:
4079466
Journal Information:
Rev. Sci. Instr., Journal Name: Rev. Sci. Instr. Vol. Vol: 35
Country of Publication:
Country unknown/Code not available
Language: