APPARATUS AND ANALYSIS FOR SUM-COINCIDENCE (n,$gamma$) SPECTROMETER
Journal Article
·
· Nuclear Instr. & Methods
Apparatus is described for analyzing two-step cascades of neutron capture gamma rays by the sum-coincidence method and consideration is given to the types of results which can be obtained. A detalled analysis is given of detection efficiency and peak shapes and their dependence on the characteristics of the sum discriminator. Several background effects are analyzed quantitatively such as back-scattering between scintlllators, annihilation photon transfer between scintillators, cascades which terminate only near to the ground state, three-photon cascades. and uncertainties caused by unknown angular correlations. (auth)
- Research Organization:
- Yale Univ., New Haven
- NSA Number:
- NSA-15-006153
- OSTI ID:
- 4076871
- Journal Information:
- Nuclear Instr. & Methods, Journal Name: Nuclear Instr. & Methods Vol. Vol: 9
- Country of Publication:
- Country unknown/Code not available
- Language:
- English