High-efficiency multidetector system for tumor scanning

Conference · · IEEE Trans. Nucl. Sci., v. NS-23, no. 1, pp. 543-548
OSTI ID:4071601

A high-efficiency detector system developed especially for medical imaging has three specially cut Ge(Li) coaxial detectors (total volume 249 cm$sup 3$). At 122 keV, the peak efficiency is 93 percent of that of a 7.6 x 7.6 cm NaI (Tl) detector. Degradation of the paralleled energy resolution is avoided and resolution is improved by 35 percent over that of conventional output-summing techniques by gating the detector outputs. In effect this multiplexes them to a single line output. (auth)

Research Organization:
Univ. of California, Livermore
NSA Number:
NSA-33-029270
OSTI ID:
4071601
Journal Information:
IEEE Trans. Nucl. Sci., v. NS-23, no. 1, pp. 543-548, Journal Name: IEEE Trans. Nucl. Sci., v. NS-23, no. 1, pp. 543-548; ISSN IETNA
Country of Publication:
United States
Language:
English