THE ANALYSIS OF URINE SPECIMENS FOR URANIUM AND PLUTONIUM
Technical Report
·
OSTI ID:4062538
Methods for analysis of urine for plutonium and uranium were studied for the purpose of recommending the most suithble methods. The sensitivity limits set were 10 dpm/ liter of urine for plutonimn and U/sup 235/ and 10 mu g/liter of urine for natural uranium. Liquid scintillation counting techniques were found unsatisfactory, and electrodepositionthan the required sensitivity, are unnecessarily complex. Alpha proportional counting in a flow counter is recommended. The recommended extraction procedures are the bismuth phosphate- lanthanum fluoride coprecipitation technique for plutonium and either the di-n- butyl orthophosphoric acid (DBP) extraction procedure or an anion excbange procedure for uranium. (auth)
- Research Organization:
- Lovelace Foundation for Medical Education and Research, Albuquerque, N. Mex.
- NSA Number:
- NSA-15-012833
- OSTI ID:
- 4062538
- Report Number(s):
- AFSWC-TN-61-8
- Country of Publication:
- United States
- Language:
- English