Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

METHOD FOR DETERMINING THE PHOTOPEAK EFFICIENCY OF SCINTILLATION COUNTERS

Journal Article · · Rev. Sci. Instr.
DOI:https://doi.org/10.1063/1.1717315· OSTI ID:4060984
A technique for determining the photopeak efficiency of certain scintillation counters for gamma rays, by detecting coincident radiations with a single Nal(Tl) counter, is described. The method is based on the fact that, for coincident radiation, a detecting system presenting 4 pi steradians to the source has an efficiency epsilon ( gamma /sub 1/) for detecting a quantum gamma /sub 1/ and an efficiency epsilon ( gamma /sub 1/) epsilon ( gamma /sub 2/) for detecting both quanta simultaneously. (M.C.G.)
Research Organization:
Texas Nuclear Corp., Austin
Sponsoring Organization:
USDOE
NSA Number:
NSA-15-014493
OSTI ID:
4060984
Journal Information:
Rev. Sci. Instr., Journal Name: Rev. Sci. Instr. Vol. Vol: 32
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

PHOTOPEAK COUNTING EFFICIENCIES FOR 3 X 3 INCH SOLID AND WELL-TYPE NaI SCINTILLATION CRYSTALS
Journal Article · Fri Sep 01 00:00:00 EDT 1961 · Analytical Chemistry (U.S.) Formerly Ind. Eng. Chem., Anal. Ed. · OSTI ID:4837455

PHOTOPEAK EFFICIENCY AND RESPONSE FUNCTION OF VARIOUS NaI(Tl) AND CsI(Tl) CRYSTALS IN THE ENERGY RANGE UP TO 11 Mev
Journal Article · Fri Nov 30 23:00:00 EST 1962 · Nucl. Instr. Methods · OSTI ID:4729582

APPARATUS FOR ROUTINE QUANTITATIVE ESTIMATION OF RADIONUCLIDES BY $gamma$- SCINTILLATION SPECTROMETRY
Technical Report · Fri Oct 31 23:00:00 EST 1958 · OSTI ID:4241473