THE SPEED OF RESPONSE OF THE 56 AVP PHOTOMULTIPLIER AND ITS MEASUREMENT IN THE LABORATORY (in French)

Journal Article · · Onde elec.
OSTI ID:4034371
S>In order to profit fully from the extremely brief duration of scintillations in time measurements in the field of nuclear physics, it is necessary to have available high-grain and high-speed photomultipliers. The Dario 56 AVP was designed with tais purpose in mind and the study of its speed of response is detailed. Special methods have been developed and the corresponding control equipment has been manufactured to measure the isochronous defects in the electron paths between the vsnious electrodes. The values calculated by the manufacturer are measured and consist of spread center to edge of the order of 0.5 ns, vaniation of total transit time in the tube less than 0.5 ns, and amplitude at half height of an anode pulse corresponding to a single photoelectron of the order of 2 ns. (auth)
Research Organization:
Centre d'Etudes Nucleaires, Saclay, France
NSA Number:
NSA-15-023706
OSTI ID:
4034371
Journal Information:
Onde elec., Journal Name: Onde elec. Vol. Vol: No. 407
Country of Publication:
Country unknown/Code not available
Language:
French

Similar Records

APPARATUS PERMITTING THE ANALYSIS OF THE PULSE OF A FAST PHOTOMULTIPLIER
Journal Article · Tue Oct 31 23:00:00 EST 1961 · Journal de Physique et le Radium (France) Changed to J. Phys. (Orsay, Fr.) · OSTI ID:4802212

Evaluation of the Amperex 56 TVP photomultiplier
Technical Report · Wed Jul 07 00:00:00 EDT 1976 · OSTI ID:7342898

Study of High Speed Photomultiplier Systems
Journal Article · Tue Dec 31 23:00:00 EST 1963 · Review of Scientific Instruments · OSTI ID:4078048