APPLICATION OF HIGH RESOLUTION LITHIUM-DRIFT GERMANIUM GAMMA-RAY SPECTROMETERS TO HIGH ENERGY GAMMA-RAYS

Journal Article · · Nucl. Instr. Methods
A description is given of the use of Li-drifted Ge detectors for high- resolution gamma spectroscopy. The detectors are used in the following manner: the photon enters the detector and annihilates into an electron-positron pair; the annihilation quanta escape from the detector; and the total energy of the electron-positron pair is lost in the detector. The detector thus reconds a peak at E -- 2m/sub 0/c/sup 2/, where E is the photon energy and mo is the electron rest mass. The method is useful only at high photon energies. The detectors in this application have energy resolutions of about 0.13%. (T.F.H.)
Research Organization:
Atomic Energy of Canada, Ltd., Chalk River, Ont.
NSA Number:
NSA-18-020374
OSTI ID:
4016116
Journal Information:
Nucl. Instr. Methods, Journal Name: Nucl. Instr. Methods Vol. Vol: 26
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

A HIGH-RESOLUTION LITHIUM-DRIFT GERMANIUM GAMMA-RAY SPECTROMETER
Journal Article · Sat Nov 30 23:00:00 EST 1963 · Nucl. Instr. Methods · OSTI ID:4053092

An observation of the Galactic center region with the HEXAGONE high resolution gamma-ray spectrometer. [Sgr A; 1742-289]
Conference · Thu Aug 01 00:00:00 EDT 1991 · AIP Conference Proceedings (American Institute of Physics); (United States) · OSTI ID:7302460

Gamma Scanning with a Lithium-Drifted Germanium Anticoincidence Spectrometer
Technical Report · Wed Aug 25 00:00:00 EDT 1965 · OSTI ID:12683203