Electron beam treatment-planning system
Some of the physical parameters necessary in electron beam treatment planning are investigated for 5- to 45-MeV electrons extracted from a Brown-- Boveri betatron. The percent depth ionization (PDI) curves are compared with the empirical relation given by Laughlin's equation, and an extension of Laughlin's equation is described. The modified absorption coefficient (MAC) method is introduced to correct the isodose distributions for the presence of inhomogeneities, such as lung, with the actual density of lung and the location of the inhomogeneity taken into consideration. Experimental data from measurements made in a water-cork phantom are presented, and the results are compared with the various calculated methods. (AIP)
- Research Organization:
- Radiation Therapy Department, Dartmouth-Hitchcock Medical Center, Hanover, New Hampshire 03755
- Sponsoring Organization:
- USDOE
- NSA Number:
- NSA-33-026955
- OSTI ID:
- 4009701
- Journal Information:
- Med. Phys., v. 3, no. 1, pp. 31-38, Other Information: Orig. Receipt Date: 30-JUN-76
- Country of Publication:
- United States
- Language:
- English
Similar Records
Treatment verification in the presence of inhomogeneities using EPID-based three-dimensional dose reconstruction
SU-E-T-219: Comprehensive Validation of the Electron Monte Carlo Dose Calculation Algorithm in RayStation Treatment Planning System for An Elekta Linear Accelerator with AgilityTM Treatment Head
Related Subjects
N42200 -Engineering-Facilities & Equipment
430303* -Particle Accelerators-Auxiliaries & Components- Experimental Facilities & Equipment
*BETATRONS- IRRADIATION DEVICES
*ELECTRON BEAMS- ISODOSE CURVES
*LUNGS- ELECTRON DOSIMETRY
MEV RANGE 10-100
PHANTOMS
RADIOTHERAPY