A method for the in vitro testing of cardiac ablation catheters

Journal Article · · IEEE Transactions on Microwave Theory and Techniques
DOI:https://doi.org/10.1109/22.539942· OSTI ID:398141
; ; ; ; ;  [1]
  1. Jefferson Medical Coll., Philadelphia, PA (United States)
The authors have developed a flow-phantom model in order to measure the temperature profile of radio frequency (RF) and microwave (MW) catheters. The model consists of a muscle equivalent phantom in a perfusion chamber with constant saline infusion of 4 L/min immersed in a 37 C saline bath. RF (4 or 8 mm, 550 kHz) or MW (12 mm helical antenna, 915 MHz) catheters were placed on the surface of the phantom and various energies were applied. Temperature measurements were obtained with fiberoptic thermometry probes placed at various distances from the catheter. Temperature contours were generated, and lesion volumes were estimated using 47 C isotherm ({Delta}T {ge} 10 C). The dosimetry of power versus {Delta}T was linear. A 2.59 fold increase in power density was required to achieve a similar surface temperature with the 8 mm versus 4 mm electrode tip. The volumes of lesions created with an 8 mm electrode were 2.5x larger than those made with a 4 mm electrode at a similar surface temperature. The RF phantom data compared favorably with the lesion volumes seen in the in vivo canine left ventricular model. The temperature profile of the microwave electrode showed heating along the length of the catheter due to imperfect tuning of the antenna. Deeper heating was seen with 8 mm RF and MW electrodes than with an RF 4 mm electrode given the same surface temperature. Measurements obtained with both a static and flow-phantom model demonstrated the cooling effects of flow on surface temperature measured during power delivery. The flow-phantom model accurately predicts the lesion geometry but underestimates the lesion volume at higher temperature when compared to the in vivo left ventricular canine model. Static phantom models will overestimate lesion size due to the surface cooling effects of cardiac blood flow. Changes in microwave catheter design may be carefully analyzed with the flow-phantom model prior to in vivo testing.
OSTI ID:
398141
Journal Information:
IEEE Transactions on Microwave Theory and Techniques, Journal Name: IEEE Transactions on Microwave Theory and Techniques Journal Issue: 10Pt2 Vol. 44; ISSN IETMAB; ISSN 0018-9480
Country of Publication:
United States
Language:
English

Similar Records

Effects of multipolar electrode radiofrequency energy delivery on ventricular endocardium
Journal Article · Wed Feb 28 23:00:00 EST 1990 · American Heart Journal; (USA) · OSTI ID:6967289

Electrode radius predicts lesion radius during radiofrequency energy heating. Validation of a proposed thermodynamic model
Journal Article · Sun Jul 01 00:00:00 EDT 1990 · Circulation Research (An Official Journal of the American Heart Association); (USA) · OSTI ID:6643284

Transcatheter desiccation of the canine left ventricle using radiofrequency energy: a pilot study
Journal Article · Wed Jul 01 00:00:00 EDT 1987 · Am. Heart J.; (United States) · OSTI ID:6320819