Carbon determination by the complex method of autoradiography-X-ray microanalysis

Journal Article · · Industrial Laboratory
OSTI ID:391311
;  [1]
  1. V.I. Vernadskii Institute of Geochemistry and Analytical Chemistry, Moscow (Russian Federation)
An x-ray microanalyzer is used for studying autoradiograms by the autoradiography - x-ray microanalysis (ARG - XRMA) method. The operating conditions of the microanalyzer are optimized by optimizing the electron accelerating voltage (15 kV), probe current (30 nA), and time of pulse counting (100 sec). X-ray background components are determined. The correctness of the results obtained is checked experimentally on glasses of basalt composition.
OSTI ID:
391311
Journal Information:
Industrial Laboratory, Journal Name: Industrial Laboratory Journal Issue: 8 Vol. 60; ISSN 0019-8447; ISSN INDLAP
Country of Publication:
United States
Language:
English

Similar Records

Cytochemical applications of X-ray microanalysis
Journal Article · Sun May 01 00:00:00 EDT 1988 · J. Electron Micros. Tech.; (United States) · OSTI ID:6611450

Advanced mercuric iodide detectors for X-ray microanalysis
Journal Article · Wed Dec 31 23:00:00 EST 1986 · Scanning Microsc. Suppl.; (United States) · OSTI ID:5384506

X-RAY MICROANALYSIS OF THE LIGHT ELEMENTS