Beamline diagnostics for protein crystallography using synchrotron radiation
- HASYLAB/DESY, D-22603 Hamburg, Germany MPG-ASMB c/o DESY, D-22603 Hamburg (Germany)
Beam diagnostics devices are described for the monitoring of all basic parameters relevant to protein data collection using synchrotron radiation. The parameters include the total power in the white beam, the positions and directions of the incident white and monochromatic beams, and the intensity, wavelength, and polarization in the incident monochromatic beam. Such devices have been installed on a doubly focusing wiggler beamline, BW6, at DORIS. They assure high quality in protein data collection despite fluctuations in the beam, facilitate rapid alignment of the entire beamline, and provide a basis for automatic realignment of the diffraction setup and the x-ray optics. {copyright} {ital 1996 American Institute of Physics.}
- OSTI ID:
- 389600
- Report Number(s):
- CONF-9510119-; ISSN 0034-6748; TRN: 96:028014
- Journal Information:
- Review of Scientific Instruments, Vol. 67, Issue 9; Conference: SRI `95: synchrotron radiation instrumentation symposium and the 7. users meeting for the advanced photon source (APS), Argonne, IL (United States), 16-20 Oct 1995; Other Information: PBD: Sep 1996
- Country of Publication:
- United States
- Language:
- English
Similar Records
Bent Diamond Crystals and Multilayer Based Optics at the new 5-Station Protein Crystallography Beamline 'Cassiopeia' at MAX-lab
New Beamlines For Protein Crystallography At The EMBL-Outstation Hamburg