Performance of Advanced Photon Source insertion devices at high photon energies (50{endash}300 keV)

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.1146617· OSTI ID:389526
; ; ;  [1]
  1. Experimental Facilities Division, Advanced Photon Source, Argonne National Laboratory, Argonne, Illinois 60439 (United States)
Since high-energy photons ({approx_gt}50 keV) are well suited for certain types of x-ray scattering experiments, we present calculated results for the Advanced Photon Source (APS) Undulator A and APS Wiggler A at high energies. The undulator calculations include the effect of magnetic field errors, which is to smear the high-order spectral harmonics. At their anticipated initial minimum gap settings, Undulator A should perform better than Wiggler A from the point of view of most high energy experiments up to at least {approximately}280 keV. A comparison of APS insertion devices to high energy insertion devices in other synchrotron radiation laboratories is also provided. {copyright} {ital 1996 American Institute of Physics.}
Research Organization:
Argonne National Laboratory (ANL), Argonne, IL
DOE Contract Number:
W-31109-ENG-38
OSTI ID:
389526
Report Number(s):
CONF-9510119--
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 9 Vol. 67; ISSN RSINAK; ISSN 0034-6748
Country of Publication:
United States