Design considerations and initial performance of a 1.2 cm{sup 2} beta imaging intra-operative probe

Journal Article · · IEEE Transactions on Nuclear Science
DOI:https://doi.org/10.1109/23.531898· OSTI ID:379773
; ; ; ;  [1]
  1. Univ. of California, Los Angeles, CA (United States)

A novel small area beta ({beta}{sup {+-}}) detector is under development for nuclear emission imaging of surgically exposed, radiolabeled tumor beds. The imaging device front-end consists of a 0.5 mm thick by 1.25 cm diameter CaF{sub 2}(Eu) scintillator disk coupled to a rapid bundle of 2 mm diameter double clad optical fibers through a polystyrene light diffuser. The detector area (1.2 cm{sup 2}) was determined by the requirement of introducing the probe into small cavities, e.g. during neuro-surgical lesion resection, but large enough to produce images of clinical significance. Flexible back-end optical fibers (1.9 m long) were coupled to the front-end components allowing {approximately} 75 photoelectrons to e detected for mean beta energies of 250 keV, indicating that sufficient signal can be obtained with clinical beta emitters (e.g. {sup 18}F, {sup 131}I). The long flexible fibers guide the scintillation light to a Philips XP1700 series, fiber optical faceplate, Multi-Channel PMT. The parallel MC-PMT outputs re fed into a variable gain, charge divider network and an i-V pre-amplifier/line driver network, whose resulting four outputs are digitized and histogrammed with standard Anger positioning logic. The various components in the imaging chain were evaluated and optimized by both simulations and measurements. Line spread functions measured in the 10.8 mm FOV were 0.50 mm {+-} 0.038 mm and 0.55 mm {+-} 0.065 mm FWHM in X and Y, respectively. A 20% variation in pulse height and minimal variation in spatial resolution was observed. The differential image uniformity was measured to be {+-}15.6% with {approximately} 150 cts/pixel. Preliminary images show excellent reproduction of phantom activity distributions.

DOE Contract Number:
FC03-87ER60615
OSTI ID:
379773
Report Number(s):
CONF-951073--
Journal Information:
IEEE Transactions on Nuclear Science, Journal Name: IEEE Transactions on Nuclear Science Journal Issue: 4Pt1 Vol. 43; ISSN 0018-9499; ISSN IETNAE
Country of Publication:
United States
Language:
English

Similar Records

Optimizing light collection from thin scintillators used in a beta-ray camera for surgical use
Journal Article · Sat Jun 01 00:00:00 EDT 1996 · IEEE Transactions on Nuclear Science · OSTI ID:267978

Characterization of fluor concentration and geometry in organic scintillators for in situ beta imaging
Journal Article · Sat Nov 30 23:00:00 EST 1996 · IEEE Transactions on Nuclear Science · OSTI ID:458243

Multi-PSPMT scintillation camera
Journal Article · Tue Jun 01 00:00:00 EDT 1999 · IEEE Transactions on Nuclear Science · OSTI ID:684505