A least squares technique for reconstructing truncated data is presented. The method involves including the ramp filter into the system matrix. The sinogram is extrapolated with zeros for filtering but extrapolated values are not considered during backward and forward projections.The new matrix leads to a frequency weighted least squares criterion. Tikhonov regularization and a priori information are considered. For the 0-order regularization, the resulting images are biased but it is shown that this bias is recovered by a simple scaling process. First and second order regularizations provide true restoration. The technique is applied to simulated and real transmission SPECT data. Results show that the frequency least squares criterion is a valuable approach for efficiently reconstructing truncated sinograms.
OSTI ID:
379769
Report Number(s):
CONF-951073--
Journal Information:
IEEE Transactions on Nuclear Science, Journal Name: IEEE Transactions on Nuclear Science Journal Issue: 4Pt1 Vol. 43; ISSN 0018-9499; ISSN IETNAE
Country of Publication:
United States
Language:
English

Similar Records

The use of the least-square minimal residual algorithm for fast and regularized attenuation compensation in SPECT
Conference · Mon Dec 30 23:00:00 EST 1996 · OSTI ID:513271

Constrained least squares filtering in high resolution PET and SPECT imaging
Conference · Sat Mar 31 23:00:00 EST 1990 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA) · OSTI ID:6871944

Digital protection of power transformer based on weighted least square algorithm
Journal Article · Sun Oct 31 23:00:00 EST 1982 · IEEE Trans. Power Appar. Syst.; (United States) · OSTI ID:6003719