Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

{sup 3}He melting pressure temperature scale below 25 mK

Journal Article · · Bulletin of the American Physical Society
OSTI ID:374895
Using {sup 60}Co {gamma} ray anisotropy radiation as a primary thermometer, with a Pt NMR susceptibility secondary thermometer, the authors have made high precision measurements of the {sup 3}He melting pressure versus temperature from 500 {mu}K to 25 mK. Temperatures obtained for the fixed points on the melting curve are: the superfluid A transition T{sub A} = 2.505 mK, the A-B transition T{sub AB} = 1.948 mK, and the solid ordering temperature T{sub N} = 0.934 mK. The authors provide a functional form for P(T), which, with the fixed points, constitutes a convenient temperature scale, based on a primary thermometer, usable to well below 1 mK.
OSTI ID:
374895
Report Number(s):
CONF-9304297--
Journal Information:
Bulletin of the American Physical Society, Journal Name: Bulletin of the American Physical Society Journal Issue: 2 Vol. 40; ISSN 0003-0503; ISSN BAPSA6
Country of Publication:
United States
Language:
English

Similar Records

{sup 3}He melting pressure thermometry
Journal Article · Sun Oct 01 00:00:00 EDT 1995 · Journal of Low Temperature Physics · OSTI ID:198186

The specific heat capacity and thermal conductivity of normal liquid /sup 3/He
Journal Article · Tue Jul 01 00:00:00 EDT 1986 · J. Low Temp. Phys.; (United States) · OSTI ID:6902171

Experiments with powdered CMN thermometers between 10 mK and 4K, and a comparison with an NBS SRM 768 fixed-point device