Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

A method for correcting the depth-of-interaction blurring in PET cameras

Journal Article · · IEEE Transactions on Medical Imaging
DOI:https://doi.org/10.1109/42.370411· OSTI ID:37339
 [1]
  1. TRIUMF, Vancouver, British Columbia (Canada)
A method is presented for correcting PET images for the blurring caused by variations in the depth-of-interaction in position-sensitive gamma ray detectors. The method uses an empirically determined, tabulated relationship between depth-of-interaction and most probable pulse-height to estimate the unknown depth from the measured pulse-height for each detected gamma ray. In the case of one fine-cut 50 x 50 x 30 mm BGO block detector, the method is shown to improve the detector resolution by 25%, averaged over the 50 x 50 mm face, measured in the geometry corresponding to detection at the edge of the field-of-view. Strengths and weaknesses of the method are discussed and its potential usefulness for improving the images of future PET cameras is assessed.
OSTI ID:
37339
Journal Information:
IEEE Transactions on Medical Imaging, Journal Name: IEEE Transactions on Medical Imaging Journal Issue: 1 Vol. 14; ISSN 0278-0062; ISSN ITMID4
Country of Publication:
United States
Language:
English

Similar Records

A practical block detector for a depth encoding PET camera
Conference · Sun Oct 01 00:00:00 EDT 1995 · OSTI ID:383078

A practical block detector for a depth-encoding PET camera
Journal Article · Sat Nov 30 23:00:00 EST 1996 · IEEE Transactions on Nuclear Science · OSTI ID:458231

Design studies of a depth encoding large aperture PET camera
Journal Article · Tue Aug 01 00:00:00 EDT 1995 · IEEE Transactions on Nuclear Science · OSTI ID:148046