Video-rate confocal scanning laser microscope for imaging human tissues {ital in vivo}

Journal Article · · Applied Optics
; ;  [1]
  1. Department of Dermatology, Wellman Laboratories, Massachusetts General Hospital, Bartlett Hall Ext. 630, 50 Blossom Street, Boston, Massachusetts 02114 (United States)

We have built a video-rate confocal scanning laser microscope for reflectance imaging of human skin and oral mucosa {ital in vivo}. Design and imaging parameters were determined for optimum resolution and contrast. Mechanical skin-holding fixtures and oral tissue clamps were made for stable objective lens-to-tissue contact such that gross tissue motion relative to the microscope was minimized. Confocal imaging was possible to maximum depths of 350 {mu}m in human skin and 450 {mu}m in oral mucosa, with measured lateral resolution of 0.5{endash}1 {mu}m and axial resolution (section thickness) of 3{endash}5 {mu}m at the 1064-nm wavelength. This resolution is comparable with that of conventional microscopy of excised biopsies (histology). Normal and abnormal tissue morphology and dynamic processes were observed. {copyright} 1999 Optical Society of America

DOE Contract Number:
FG02-91ER61229
OSTI ID:
341348
Journal Information:
Applied Optics, Journal Name: Applied Optics Journal Issue: 10 Vol. 38; ISSN APOPAI; ISSN 0003-6935
Country of Publication:
United States
Language:
English

Similar Records

Confocal microscope with large field and working distance
Journal Article · Sun Aug 01 00:00:00 EDT 1999 · Applied Optics · OSTI ID:690672

Enhancing Diagnosis through AI-driven Analysis of Reflectance Confocal Microscopy
Conference · Mon Apr 01 00:00:00 EDT 2024 · OSTI ID:2351058

Spectrally encoded confocal microscopy
Journal Article · Sat Aug 01 00:00:00 EDT 1998 · Optics Letters · OSTI ID:639074