Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Modular 64x64 CdZnTe arrays with multiplexer readout for high-resolution nuclear medicine imaging

Book ·
OSTI ID:323851
; ; ; ; ; ; ; ;  [1];  [2]
  1. Univ. of Arizona, Tucson, AZ (United States)
  2. Augustine Engineering, Encinitas, CA (United States)

The authors are developing modular arrays of CdZnTe radiation detectors for high-resolution nuclear medicine imaging. Each detector is delineated into a 64x64 array of pixels; the pixel pitch is 380 {micro}m. Each pixel is connected to a corresponding pad on a multiplexer readout circuit. The imaging system is controlled by a personal computer. They obtained images of standard nuclear medicine phantoms in which the spatial resolution of approximately 1.5 mm was limited by the collimator that was used. Significant improvements in spatial resolution should be possible with different collimator designs. These results are promising for high-resolution nuclear medicine imaging.

Sponsoring Organization:
National Cancer Inst., Bethesda, MD (United States)
OSTI ID:
323851
Report Number(s):
CONF-971201--
Country of Publication:
United States
Language:
English

Similar Records

Progress in the development of large-area modular 64 x 64 CdZnTe imaging arrays for nuclear medicine
Journal Article · Mon Jun 01 00:00:00 EDT 1998 · IEEE Transactions on Nuclear Science · OSTI ID:624148

A 48x48 CdZnTe array with multiplexer readout
Journal Article · Sat Jun 01 00:00:00 EDT 1996 · IEEE Transactions on Nuclear Science · OSTI ID:277648

CdZnTe arrays for nuclear medicine imaging
Conference · Mon Dec 30 23:00:00 EST 1996 · OSTI ID:524735