We scanned VULCAN chopper by different wavelength band center and band width. Three commonly used settings 60, 30, and 20 Hz are investigated for existing detector banks. The full scanned results in wavelength unit are included in this report. From the result, one can decide what chopper setting can be used to meet experiment requirement while avoiding time-of-flight frame overlapping or leakage, which generate artifacts in collected neutron data.
Research Organization:
Oak Ridge National Laboratory (ORNL), Oak Ridge, TN (United States)
Sponsoring Organization:
USDOE Office of Science (SC), Basic Energy Sciences (BES)
DOE Contract Number:
AC05-00OR22725
OSTI ID:
3007858
Report Number(s):
ORNL/TM--2025-4350
Country of Publication:
United States
Language:
English

Similar Records

Analysis of a drum chopper for use on a new small angle diffractometer at IPNS
Conference · Sun Dec 31 23:00:00 EST 1989 · OSTI ID:6147856

Wavelength modulation spectroscopy using novel mechanical light chopper blade designs
Journal Article · Mon Aug 15 00:00:00 EDT 2005 · Review of Scientific Instruments · OSTI ID:20723099

Multiple-angle time-of-flight design concepts. Are we overlooking an order of magnitude in instrument performance for general structural studies