Far infrared tangential interferometry/polarimetry on the National Spherical Tokamak Experiment
- Princeton Plasma Physics Laboratory, Princeton, New Jersey 08543 (United States)
- Department of Applied Science, University of California at Davis, California 95616 (United States)
Measurement of the core B{sub T}(r,t) value is essential in the National Spherical Tokamak Experiment (NSTX), since the effects of paramagnetism and diamagnetism in the NSTX are expected to be considerably greater than that in higher aspect ratio tokamaks. Therefore, without independent B{sub T}(r,t) measurement, plasma parameters dependent upon B{sub T} such as the {ital q} profile and local {beta} value cannot be evaluated. Tangential interferometer/polarimeter systems (eight channels) [H. Park, L. Guttadora, C. Domier, W. R. Geck, and N. C. Luhman, Jr., First and Second NSTX Research Forums, Princeton, NJ, 1997 (unpublished)] for the NSTX will provide temporally and radially resolved toroidal field profile [B{sub T}(r,t)] and two-dimensional electron density profile [n{sub e}(r,t)] data. The outcome of the proposed system is extremely important to the study of confinement, heating, and stability of the NSTX plasmas. The research task is largely based on utilizing existing hardware from the TFTR multichannel infrared interferometer system [D. K. Mansfield, H. K. Park, L. C. Johnson, H. Anderson, S. Foote, B. Clifton, and C. H. Ma, Appl. Opt. {bold 26}, 4469 (1987) and H. K. Park, D. K. Mansfield, and C. L. Johnson, Proceedings of the 3rd International Symposium on Laser-Aided Plasma Diagnostic, Los Angeles, CA, 28{endash}30 Oct. 1987 (unpublished), pp. 96{endash}104] which will be reconfigured into a tangential system for NSTX, and to develop the additional hardware required to complete the system. {copyright} {ital 1999 American Institute of Physics.}
- OSTI ID:
- 295647
- Report Number(s):
- CONF-980605--
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 1 Vol. 70; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
Status of Far Infrared Tangential Interferometry/Polarimetry (FIReTIP) on NSTX
Application of Stark Tuned Laser for Interferometry and Polarimetry in Plasmas