Real-time system for respiratory-cardiac gating in positron tomography

Journal Article · · IEEE Transactions on Nuclear Science
DOI:https://doi.org/10.1109/23.708323· OSTI ID:292928
; ; ;  [1];  [2]
  1. Lawrence Berkeley National Lab., CA (United States). Center for Functional Imaging
  2. CTI, Inc., Knoxville, TN (United States)
A Macintosh-based signal processing system has been developed to support simultaneous respiratory and cardiac gating on the ECAT EXACT HR PET scanner. Using the Lab-View real-time software environment, the system reads analog inputs from a pneumatic respiratory bellows and an EGG monitor to compute an appropriate histogram memory location for the PET data. Respiratory state is determined by the bellows signal amplitude; cardiac state is based on the time since the last R-wave. These two states are used in a 2D lookup table to determine a combined respiratory-cardiac state. A 4-bit address encoding the selected histogram is directed from the system to the ECAT scanner, which dynamically switches the destination of tomograph events as respiratory-cardiac state changes. to Test the switching efficiency of the combined Macintosh/ECAT system, a rotating emission phantom was built. Acquisitions with 25 msec states while the phantom was rotating at 240 rpm demonstrate the system could effectively stop motion at this rate, with approximately 5 msec switching time between states.
Sponsoring Organization:
Department of Health and Human Services, Washington, DC (United States); USDOE Office of Energy Research, Washington, DC (United States)
DOE Contract Number:
AC03-76SF00098
OSTI ID:
292928
Report Number(s):
CONF-971147--
Journal Information:
IEEE Transactions on Nuclear Science, Journal Name: IEEE Transactions on Nuclear Science Journal Issue: 4Pt2 Vol. 45; ISSN 0018-9499; ISSN IETNAE
Country of Publication:
United States
Language:
English

Similar Records

Acquisition and automated 3-D segmentation of respiratory/cardiac-gated PET transmission images
Conference · Mon Dec 30 23:00:00 EST 1996 · OSTI ID:513210

Respiratory compensation in cardiac PET using doubly-gated acquisitions
Journal Article · Thu May 01 00:00:00 EDT 1997 · Journal of Nuclear Medicine · OSTI ID:563341

Compensation for respiratory motion in cardiac PET - A feasibility study
Journal Article · Wed May 01 00:00:00 EDT 1996 · Journal of Nuclear Medicine · OSTI ID:447768