Absorbed dose measurements and calculations in phantoms for 1.5 to 50 keV photons

Journal Article · · Health Physics
 [1]; ;  [2]
  1. Ship Research Institute, Tokyo (Japan)
  2. Japan Atomic Energy Research Institute, Ibaraki (Japan); and others

A Monte Carlo code EGS4 expanded for low energy photon transport was validated by measuring absorbed doses in a phantom for 30 and 10 keV monoenergetic photons from synchrotron radiation. Using the EGS4 code, depth doses at 0.07 mm, 0.02 to 0.1 mm, and 10 mm in the ICRU slab phantoms were calculated for 1.5 to 50 keV photons using the updated photon cross section data PHOTX. The results show that the doses at 0.02 to 0.1 mm below 10 keV are practical indices of effective dose as calculated by others. 19 refs., 7 figs., 7 tabs.