Title: Evaluation of moderately cooled pure NaI as a scintillator for position-sensitive PET detectors

Journal Article · · IEEE Transactions on Nuclear Science
DOI:https://doi.org/10.1109/23.507251· OSTI ID:276524
; ; ;  [1]
  1. Univ. of Pennsylvania, Philadelphia, PA (United States). Dept. of Radiology

A new evaluation of pure NaI has been performed to determine if moderate cooling would lead to better performance than that of existing, activated NaI(Tl) position-sensitive detectors, particularly at high countrates. Using a freezer, an initial effort was performed to cool the crystal assembly to {minus}90 C (183 K). At this temperature, pure NaI has a decay constant of 35 nsec, a light output which is about 20% that of room temperature NaI(Tl), and an energy resolution of 15%. For the PET applications the signal of room temperature (25 C) NaI(Tl) is normally pulse clipped, reducing the light output to 40% of the unclipped signal and yielding an energy resolution of 10.5%. Since the long decay of NaI(Tl) causes it to suffer more significantly than pure NaI from pre-pulse pileup, the difference in energy resolution between the two crystals at high countrates will be reduced. Also, a significantly shorter trigger deadtime with pure NaI will lead to a reduction in coincidence deadtime losses in PET. Computer simulations of large-area crystals operating at high countrates have been performed to quantify their trigger deadtime behavior and position resolution as a function of light output and pulse decay time. Having gained experience with the practical issues of cooling large crystals, measurements of position resolution have been performed with a NaI bar detector of similar geometry to the NaI(Tl) detectors in use in the PENN-PET scanner.

DOE Contract Number:
FG02-88ER60642
OSTI ID:
276524
Report Number(s):
CONF-951073-; ISSN 0018-9499; TRN: 96:018029
Journal Information:
IEEE Transactions on Nuclear Science, Vol. 43, Issue 3Pt2; Conference: IEEE nuclear science symposium and medical imaging conference, San Francisco, CA (United States), 21-28 Oct 1995; Other Information: PBD: Jun 1996
Country of Publication:
United States
Language:
English

Similar Records

Comparison of multi-pole shaping and delay line clipping pre-amplifiers for position sensitive NaI(Tl) detectors
Journal Article · Mon Jun 01 00:00:00 EDT 1998 · IEEE Transactions on Nuclear Science · OSTI ID:276524

A model of the high count rate performance of NaI(Tl)-based PET detectors
Journal Article · Mon Jun 01 00:00:00 EDT 1998 · IEEE Transactions on Nuclear Science · OSTI ID:276524

Design and performance of HEAD PENN-PET scanner
Conference · Mon Aug 01 00:00:00 EDT 1994 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:276524