skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: A study of laseruler accuracy and precision (1986-1987)

Technical Report ·
DOI:https://doi.org/10.2172/266360· OSTI ID:266360

A study was conducted to investigate Laserruler accuracy and precision. Tests were performed on 0.050 in., 0.100 in., and 0.120 in. gauge block standards. Results showed and accuracy of 3.7 {mu}in. for the 0.12 in. standard, with higher accuracies for the two thinner blocks. The Laserruler precision was 4.83 {mu}in. for the 0.120 in. standard, 3.83 {mu}in. for the 0.100 in. standard, and 4.2 {mu}in. for the 0.050 in. standard.

Research Organization:
EG and G Mound Applied Technologies, Miamisburg, OH (United States)
Sponsoring Organization:
USDOE, Washington, DC (United States)
DOE Contract Number:
AC24-88DP43495
OSTI ID:
266360
Report Number(s):
MLM-MU-89-66-0001; ON: DE96013047
Resource Relation:
Other Information: PBD: 22 Jun 1989
Country of Publication:
United States
Language:
English