The updated besge spectrometer for pure beta activities in environmental matrices
Journal Article
·
· Health Physics
- Universita di Genova (Italy)
This work describes an updated version of a beta spectrometer that measures pure beta emitter activities. The detection system consists of a plastic scintillator surrounded in a air geometry by large anticoincidence NaI(Tl) scintillators. A thin silicon detector, inserted between the plastic and the sample, helps to reduce the gamma background. A Monte Carlo code, used to calculate the quite complex spectrometer response function to any isotope, is also described. 10 refs., 4 figs., 1 tab.
- Sponsoring Organization:
- USDOE
- OSTI ID:
- 258994
- Journal Information:
- Health Physics, Journal Name: Health Physics Journal Issue: 4 Vol. 66; ISSN HLTPAO; ISSN 0017-9078
- Country of Publication:
- United States
- Language:
- English