Determining the tilt of the Raman laser beam using an optical method for atom gravimeters
- Huazhong Univ. of Science and Technology, Wuhan (China)
- Huazhong Univ. of Science and Technology, Wuhan (China); Oak Ridge National Laboratory (ORNL), Oak Ridge, TN (United States)
The tilt of a Raman laser beam is a major systematic error in precision gravity measurement using atom interferometry. The conventional approach to evaluating this tilt error involves modulating the direction of the Raman laser beam and conducting time-consuming gravity measurements to identify the error minimum. In this work, we demonstrate a method to expediently determine the tilt of the Raman laser beam by transforming the tilt angle measurement into characterization of parallelism, which integrates the optical method of aligning the laser direction, commonly used in freely falling corner-cube gravimeters, into an atom gravimeter. A position-sensing detector (PSD) is utilized to quantitatively characterize the parallelism between the test beam and the reference beam, thus measuring the tilt precisely and rapidly. After carefully positioning the PSD and calibrating the relationship between the distance measured by the PSD and the tilt angle measured by the tiltmeter, we achieved a statistical uncertainty of less than 30 µrad in the tilt measurement. Furthermore, we compared the results obtained through this optical method with those from the conventional tilt modulation method for gravity measurement. The comparison validates that our optical method can achieve tilt determination with an accuracy level of better than 200 µrad, corresponding to a systematic error of 20 µGal in g measurement. This work has practical implications for real-world applications of atom gravimeters.
- Research Organization:
- Oak Ridge National Laboratory (ORNL), Oak Ridge, TN (United States)
- Sponsoring Organization:
- National Key Research and Development Program of China; USDOE
- Grant/Contract Number:
- AC05-00OR22725
- OSTI ID:
- 2478342
- Journal Information:
- Chinese Physics B, Journal Name: Chinese Physics B Vol. 33; ISSN 1674-1056
- Publisher:
- IOP PublishingCopyright Statement
- Country of Publication:
- United States
- Language:
- English
Similar Records
Evaluation of a frequency-dependent phase shift in chirped Raman lasers for atom gravimeters
Double diffraction in an atomic gravimeter
SYSTEMATIC ERROR REDUCTION: NON-TILTED REFERENCE BEAM METHOD FOR LONG TRACE PROFILER.
Journal Article
·
Wed Dec 18 19:00:00 EST 2024
· Physical Review A
·
OSTI ID:3002001
Double diffraction in an atomic gravimeter
Journal Article
·
Thu Jan 14 23:00:00 EST 2010
· Physical Review. A
·
OSTI ID:21388779
SYSTEMATIC ERROR REDUCTION: NON-TILTED REFERENCE BEAM METHOD FOR LONG TRACE PROFILER.
Conference
·
Sat Aug 25 00:00:00 EDT 2007
·
OSTI ID:918596