Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Monte Carlo Simulation and Reconstruction: Assessment of Myocardial Perfusion Imaging of Tracer Dynamics With Cardiac Motion Due to Deformation and Respiration Using Gamma Camera With Continuous Acquisition

Journal Article · · Frontiers in Cardiovascular Medicine
 [1];  [1];  [2];  [3]
  1. Univ. of California, San Francisco, CA (United States)
  2. Univ. of California, San Francisco, CA (United States); Lawrence Berkeley National Laboratory (LBNL), Berkeley, CA (United States)
  3. Univ. of California, San Francisco, CA (United States); Lawrence Berkeley National Laboratory (LBNL), Berkeley, CA (United States); University of California, Berkeley, CA (United States)
Purpose: Myocardial perfusion imaging (MPI) with single photon emission computed tomography (SPECT) is routinely used for stress testing in nuclear medicine. Recently, our group extended its potential going from 3D visual qualitative image analysis to 4D spatiotemporal reconstruction of dynamically acquired data to capture the time variation of the radiotracer concentration and the estimated myocardial blood flow (MBF) and coronary flow reserve (CFR). However, the quality of reconstructed image is compromised due to cardiac deformation and respiration. The work presented here develops an algorithm that reconstructs the dynamic sequence of separate respiratory and cardiac phases and evaluates the algorithm with data simulated with a Monte Carlo simulation for the continuous image acquisition and processing with a slowly rotating SPECT camera. Methods: A clinically realistic Monte Carlo (MC) simulation is developed using the 4D Extended Cardiac Torso (XCAT) digital phantom with respiratory and cardiac motion to model continuous data acquisition of dynamic cardiac SPECT with slowly rotating gamma cameras by incorporating deformation and displacement of the myocardium due to cardiac and respiratory motion. We extended our previously developed 4D maximum-likelihood expectation-maximization (MLEM) reconstruction algorithm for a data set binned from a continuous list mode (LM) simulation with cardiac and respiratory information. Our spatiotemporal image reconstruction uses splines to explicitly model the temporal change of the tracer for each cardiac and respiratory gate that delineates the myocardial spatial position as the tracer washes in and out. Unlike in a fully list-mode data acquisition and reconstruction the accumulated photons are binned over a specific but very short time interval corresponding to each cardiac and respiratory gate. Reconstruction results are presented showing the dynamics of the tracer in the myocardium as it continuously deforms. These results are then compared with the conventional 4D spatiotemporal reconstruction method that models only the temporal changes of the tracer activity. Mean Stabilized Activity (MSA), signal to noise ratio (SNR) and Bias for the myocardium activities for three different target-to-background ratios (TBRs) are evaluated. Dynamic quantitative indices such as wash-in (K1) and wash-out (k2) rates at each gate were also estimated. Results: The MSA and SNR are higher with higher TBRs while biases were improved with higher TBRs to less than 10%. The correlation between exhalation-inhalation sequence with the ground truth during respiratory cycle was excellent. Our reconstruction method showed better resolved myocardial walls during diastole to systole as compared to the ungated 4D image. Estimated values of K1 and k2 were also consistent with the ground truth. Conclusion: The continuous image acquisition for dynamic scan using conventional two-head gamma cameras can provide valuable information for MPI. Our study demonstrated the viability of using a continuous image acquisition method on a widely used clinical two-head SPECT system. Our reconstruction method showed better resolved myocardial walls during diastole to systole as compared to the ungated 4D image. Precise implementation of reconstruction algorithms, better segmentation techniques by generating images of different tissue types and background activity would improve the feasibility of the method in real clinical environment.
Research Organization:
Lawrence Berkeley National Laboratory (LBNL), Berkeley, CA (United States)
Sponsoring Organization:
National Institutes of Health (NIH); USDOE
Grant/Contract Number:
AC02-05CH11231
OSTI ID:
2470666
Journal Information:
Frontiers in Cardiovascular Medicine, Journal Name: Frontiers in Cardiovascular Medicine Vol. 9; ISSN 2297-055X
Publisher:
Frontiers Media S.A.Copyright Statement
Country of Publication:
United States
Language:
English

References (44)

Prognostic value of myocardial perfusion SPECT images in combination with the maximal heart rate at exercise testing in Japanese patients with suspected ischemic heart disease: a sub-analysis of J-ACCESS journal November 2009
SPECT Imaging for Detecting Coronary Artery Disease and Determining Prognosis by Noninvasive Assessment of Myocardial Perfusion and Myocardial Viability journal July 2011
Measurement of absolute myocardial blood flow in humans using dynamic cardiac SPECT and 99mTc-tetrofosmin: Method and validation journal December 2015
A combined static-dynamic single-dose imaging protocol to compare quantitative dynamic SPECT with static conventional SPECT journal August 2017
Single Photon Emission Computed Tomography (SPECT) Myocardial Perfusion Imaging Guidelines: Instrumentation, Acquisition, Processing, and Interpretation journal May 2018
Four-Dimensional Image Reconstruction Strategies in Cardiac-Gated and Respiratory-Gated PET Imaging journal January 2013
Technological Development and Advances in Single-Photon Emission Computed Tomography/Computed Tomography journal May 2008
A Monte Carlo investigation of dual-energy-window scatter correction for volume-of-interest quantification in99TcmSPECT journal January 1995
Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam projections journal April 1998
Rigid-body transformation of list-mode projection data for respiratory motion correction in cardiac PET journal July 2005
Regularized B-spline deformable registration for respiratory motion correction in PET images journal April 2009
Dynamic single photon emission computed tomography—basic principles and cardiac applications journal September 2010
GATE V6: a major enhancement of the GATE simulation platform enabling modelling of CT and radiotherapy journal January 2011
Investigation of dynamic SPECT measurements of the arterial input function in human subjects using simulation, phantom and human studies journal December 2011
Comparison of basis functions for 3D PET reconstruction using a Monte Carlo system matrix journal March 2012
Motion-compensated PET image reconstruction with respiratory-matched attenuation correction using two low-dose inhale and exhale CT images journal September 2013
4D image reconstruction for emission tomography journal October 2014
Image reconstruction in higher dimensions: myocardial perfusion imaging of tracer dynamics with cardiac motion due to deformation and respiration journal October 2015
Accelerated GPU based SPECT Monte Carlo simulations journal May 2016
Motion correction of PET images using multiple acquisition frames journal April 1997
Direct least-squares estimation of spatiotemporal distributions from dynamic SPECT projections using a spatial segmentation and temporal B-splines journal May 2000
Maximum Likelihood Reconstruction for Emission Tomography journal October 1982
Spatiotemporal reconstruction of list-mode PET data journal April 2002
Direct List Mode Parametric Reconstruction for Dynamic Cardiac SPECT journal January 2020
Estimation of Rigid-Body and Respiratory Motion of the Heart From Marker-Tracking Data for SPECT Motion Correction journal February 2009
Regularized Fully 5D Reconstruction of Cardiac Gated Dynamic SPECT Images journal June 2010
Incorporation of the Living Heart Model Into the 4-D XCAT Phantom for Cardiac Imaging Research journal January 2019
Reconstruction for Gated Dynamic Cardiac PET Imaging Using a Tensor Product Spline Basis journal February 2007
Four-dimensional (4D) image reconstruction strategies in dynamic PET: Beyond conventional independent frame reconstruction: Four-dimensional PET imaging journal July 2009
Respiratory motion correction in gated cardiac SPECT using quaternion-based, rigid-body registration: Quaternion-based respiratory motion correction in cardiac SPECT journal September 2009
4D XCAT phantom for multimodality imaging research: 4D XCAT phantom for multimodality imaging research journal August 2010
Fast calculation of the exact radiological path for a three-dimensional CT array: Technical Reports: 3D CT array path calculation journal March 1985
Artifacts at Cardiac CT: Physics and Solutions journal November 2016
Relation Among Stenosis Severity, Myocardial Blood Flow, and Flow Reserve in Patients With Coronary Artery Disease journal April 1995
Quantification of regional myocardial blood flow using 13N-ammonia and reoriented dynamic positron emission tomographic imaging. journal September 1992
Exercise Myocardial Perfusion SPECT in Patients Without Known Coronary Artery Disease journal March 1996
Clinical validation of time reduction strategy in continuous step-and-shoot mode during SPECT acquisition journal February 2021
Clinical evaluation of General Electric new Swiftscan solution in bone scintigraphy on NaI-camera: A head to head comparison with Siemens Symbia journal September 2019
Correction of Heart Motion Due to Respiration in Clinical Myocardial Perfusion SPECT Scans Using Respiratory Gating journal April 2007
Cardiac Dedicated Ultrafast SPECT Cameras: New Designs and Clinical Implications journal January 2011
Quantification of Myocardial Perfusion Reserve Using Dynamic SPECT Imaging in Humans: A Feasibility Study journal April 2013
Dynamic SPECT Measurement of Absolute Myocardial Blood Flow in a Porcine Model journal September 2014
Solid-State Detector SPECT Myocardial Perfusion Imaging journal August 2019
Faster Acquisition for Dopamine Transporter Imaging Using Swiftscan Step and Shoot Continuous SPECT Without Impairing Visual and Semiquantitative Analysis journal June 2020