Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

GADRAS-DRF Enhancements for Safeguards – Custom Peak Fitting to Enhance Model Fitting and Isotopics

Technical Report ·
DOI:https://doi.org/10.2172/2462959· OSTI ID:2462959
 [1];  [1]
  1. Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)
One major software update was accomplished within the Gamma Detector Response and Analysis Software-Detector Response Function (GADRAS-DRF) package. This update allows users to adjust individual peak fits for use in subsequent analysis processes within GADRAS-DRF. A graphical user interface (GUI) was implemented so users can see the effect of their fit adjustments on the spectrum. This new feature will enhance the capability of the previously funded auto-enrichment and peak-based 1D model fit feature that was implemented in FY23 (Fiscal Year). Isotopic analysis was performed using M400 data obtained from uranium standards and the isotopic assessment is given with and without manual peak adjustments.
Research Organization:
Sandia National Laboratories (SNL-NM), Albuquerque, NM (United States)
Sponsoring Organization:
USDOE National Nuclear Security Administration (NNSA)
DOE Contract Number:
NA0003525
OSTI ID:
2462959
Report Number(s):
SAND--2024-12182
Country of Publication:
United States
Language:
English

Similar Records

GADRAS-DRF Validation for Safeguards and Custom Peak Fit Enhancements
Technical Report · Mon Sep 01 00:00:00 EDT 2025 · OSTI ID:2997330

GADRAS-DRF user's manual.
Technical Report · Sun Sep 01 00:00:00 EDT 2013 · OSTI ID:1096506

GADRAS-DRF 18.6 User's Manual
Technical Report · Sun May 01 00:00:00 EDT 2016 · OSTI ID:1431293