Implementation of an impurity diagnostic suite on the Pegasus-III experiment
Journal Article
·
· Review of Scientific Instruments
- University of Wisconsin-Madison, WI (United States); University of Wisconsin-Madison
- University of Wisconsin-Madison, WI (United States)
A suite of diagnostics used to assess impurity content and dynamics has been updated, upgraded, and installed on the Pegasus-III Experiment. Typical plasma parameters during local helicity injection start-up are τshot ~10 ms, ne ~ 1 × 1019 m–3, and Te ~ 50 eV. The deployed diagnostics are compatible with this modest temperature and density regime and provide species identification, source localization, and estimation of radiation losses. Impurity species are determined by recording time-evolving, single line-of-sight spectra at 1.25 kfps using a SPRED (Survey, Poor Resolution, Extended Domain) vacuum ultraviolet spectrometer. SPRED is equipped with 450 g/mm grating, giving a spectral resolution of 0.33 nm and a spectral range from ~10 to 110 nm, useful to identify light impurity species in this temperature and density range. An absolutely calibrated spectrometer that collects light from the plasma at Rtan = 15.9 cm and Δt ≥ 2 ms is used as a visible survey spectrometer and for continuum measurements. The radiated power from the plasma is estimated with a photodiode-based diagnostic. Two 16-channel absolute extreme ultraviolet diode arrays are placed behind pinhole apertures, resulting in 32 lines of sight at Z = 0, with a spatial resolution of 2–3 cm and a time response of 60 kHz. A photometrically calibrated collinear Dα/near infrared filtered photodiode-based system measures the Dα emission and around 1040 nm. All these instruments have been designed to suppress electromagnetic interference from megawatt-class switching power supplies.
- Research Organization:
- University of Wisconsin-Madison, WI (United States)
- Sponsoring Organization:
- USDOE Office of Science (SC), Fusion Energy Sciences (FES)
- Grant/Contract Number:
- SC0019008
- OSTI ID:
- 2453877
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 8 Vol. 95; ISSN 0034-6748
- Publisher:
- American Institute of Physics (AIP)Copyright Statement
- Country of Publication:
- United States
- Language:
- English
Similar Records
Charge exchange recombination spectroscopy on the T-10 tokamak
High-resolution imaging spectrometer for recording absolutely calibrated far ultraviolet spectra from laser-produced plasmas
Journal Article
·
Sun May 15 00:00:00 EDT 2016
· Review of Scientific Instruments
·
OSTI ID:22598031
High-resolution imaging spectrometer for recording absolutely calibrated far ultraviolet spectra from laser-produced plasmas
Journal Article
·
Wed Oct 15 00:00:00 EDT 2008
· Review of Scientific Instruments
·
OSTI ID:21266445