skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: IR-MALDESI Mass Spectrometry Imaging at 50 Micron Spatial Resolution

Journal Article · · Journal of the American Society for Mass Spectrometry
 [1];  [2]; ; ;
  1. North Carolina State University, W.M. Keck FTMS Laboratory for Human Health Research, Department of Chemistry (United States)
  2. JGM Associates Inc. (United States)

High spatial resolution in mass spectrometry imaging (MSI) is crucial to understanding the biology dictated by molecular distributions in complex tissue systems. Here, we present MSI using infrared matrix-assisted laser desorption electrospray ionization (IR-MALDESI) at 50 μm resolution. An adjustable iris, beam expander, and an aspherical focusing lens were used to reduce tissue ablation diameters for MSI at high resolution. The laser beam caustic was modeled using laser ablation paper to calculate relevant laser beam characteristics. The minimum laser spot diameter on the tissue was determined using tissue staining and microscopy. Finally, the newly constructed optical system was used to image hen ovarian tissue with and without oversampling, detailing tissue features at 50 μm resolution. .

OSTI ID:
22776864
Journal Information:
Journal of the American Society for Mass Spectrometry, Vol. 28, Issue 10; Other Information: Copyright (c) 2017 American Society for Mass Spectrometry; http://www.springer-ny.com; Country of input: International Atomic Energy Agency (IAEA); ISSN 1044-0305
Country of Publication:
United States
Language:
English