Feasibility of MRI-only treatment planning for proton therapy in brain and prostate cancers: Dose calculation accuracy in substitute CT images
- Department of Radiation Oncology, Comprehensive Cancer Center, Helsinki University Central Hospital, P.O. Box 180, Helsinki 00029 HUS (Finland)
Purpose: Magnetic resonance imaging (MRI) is increasingly used for radiotherapy target delineation, image guidance, and treatment response monitoring. Recent studies have shown that an entire external x-ray radiotherapy treatment planning (RTP) workflow for brain tumor or prostate cancer patients based only on MRI reference images is feasible. This study aims to show that a MRI-only based RTP workflow is also feasible for proton beam therapy plans generated in MRI-based substitute computed tomography (sCT) images of the head and the pelvis. Methods: The sCTs were constructed for ten prostate cancer and ten brain tumor patients primarily by transforming the intensity values of in-phase MR images to Hounsfield units (HUs) with a dual model HU conversion technique to enable heterogeneous tissue representation. HU conversion models for the pelvis were adopted from previous studies, further extended in this study also for head MRI by generating anatomical site-specific conversion models (a new training data set of ten other brain patients). This study also evaluated two other types of simplified sCT: dual bulk density (for bone and water) and homogeneous (water only). For every clinical case, intensity modulated proton therapy (IMPT) plans robustly optimized in standard planning CTs were calculated in sCT for evaluation, and vice versa. Overall dose agreement was evaluated using dose–volume histogram parameters and 3D gamma criteria. Results: In heterogeneous sCTs, the mean absolute errors in HUs were 34 (soft tissues: 13, bones: 92) and 42 (soft tissues: 9, bones: 97) in the head and in the pelvis, respectively. The maximum absolute dose differences relative to CT in the brain tumor clinical target volume (CTV) were 1.4% for heterogeneous sCT, 1.8% for dual bulk sCT, and 8.9% for homogenous sCT. The corresponding maximum differences in the prostate CTV were 0.6%, 1.2%, and 3.6%, respectively. The percentages of dose points in the head and pelvis passing 1% and 1 mm gamma index criteria were over 91%, 85%, and 38% with heterogeneous, dual bulk, and homogeneous sCTs, respectively. There were no significant changes to gamma index pass rates for IMPT plans first optimized in CT and then calculated in heterogeneous sCT versus IMPT plans first optimized in heterogeneous sCT and then calculated on standard CT. Conclusions: This study demonstrates that proton therapy dose calculations on heterogeneous sCTs are in good agreement with plans generated with standard planning CT. An MRI-only based RTP workflow is feasible in IMPT for brain tumors and prostate cancers.
- OSTI ID:
- 22689404
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 8 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
A dual model HU conversion from MRI intensity values within and outside of bone segment for MRI-based radiotherapy treatment planning of prostate cancer
A dual model HU conversion from MRI intensity values within and outside of bone segment for MRI-based radiotherapy treatment planning of prostate cancer
Technical Note: MRI only prostate radiotherapy planning using the statistical decomposition algorithm
Journal Article
·
Tue Jan 14 23:00:00 EST 2014
· Medical Physics
·
OSTI ID:22251240
A dual model HU conversion from MRI intensity values within and outside of bone segment for MRI-based radiotherapy treatment planning of prostate cancer
Journal Article
·
Tue Jan 14 23:00:00 EST 2014
· Medical Physics
·
OSTI ID:22251733
Technical Note: MRI only prostate radiotherapy planning using the statistical decomposition algorithm
Journal Article
·
Thu Oct 15 00:00:00 EDT 2015
· Medical Physics
·
OSTI ID:22482367