Estimating scatter in cone beam CT with striped ratio grids: A preliminary investigation
- Department of Radiology, Stanford University, Stanford, California 94305 (United States)
Purpose: To propose a new method for estimating scatter in x-ray imaging. Conventional antiscatter grids reject scatter at an efficiency that is constant or slowly varying over the surface of the grid. A striped ratio antiscatter grid, composed of stripes that alternate between high and low grid ratio, could be used instead. Such a striped ratio grid would reduce scatter-to-primary ratio as a conventional grid would, but more importantly, the signal discontinuities at the boundaries of stripes can be used to estimate local scatter content. Methods: Signal discontinuities provide information on scatter, but are contaminated by variation in primary radiation. A nonlinear image processing algorithm is used to estimate the scatter content in the presence of primary variation. We emulated a striped ratio grid by imaging phantoms with two sequential CT scans, one with and one without a conventional grid. These two scans are processed together to mimic a striped ratio grid. This represents a best case limit of the striped ratio grid, in that the extent of grid ratio modulation is very high and the scatter contrast is maximized. Results: In a uniform cylinder, the striped ratio grid virtually eliminates cupping. Artifacts from scatter are improved in an anthropomorphic phantom. Some banding artifacts are induced by the striped ratio grid. Conclusions: Striped ratio grids could be a simple and effective evolution of conventional antiscatter grids. Construction and validation of a physical prototype remains an important future step.
- OSTI ID:
- 22689393
- Journal Information:
- Medical Physics, Vol. 43, Issue 9; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
Magnitude and effects of x-ray scatter in a 256-slice CT scanner
Improved image quality of cone beam CT scans for radiotherapy image guidance using fiber-interspaced antiscatter grid