Estimating scatter in cone beam CT with striped ratio grids: A preliminary investigation

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4960631· OSTI ID:22689393
 [1]
  1. Department of Radiology, Stanford University, Stanford, California 94305 (United States)

Purpose: To propose a new method for estimating scatter in x-ray imaging. Conventional antiscatter grids reject scatter at an efficiency that is constant or slowly varying over the surface of the grid. A striped ratio antiscatter grid, composed of stripes that alternate between high and low grid ratio, could be used instead. Such a striped ratio grid would reduce scatter-to-primary ratio as a conventional grid would, but more importantly, the signal discontinuities at the boundaries of stripes can be used to estimate local scatter content. Methods: Signal discontinuities provide information on scatter, but are contaminated by variation in primary radiation. A nonlinear image processing algorithm is used to estimate the scatter content in the presence of primary variation. We emulated a striped ratio grid by imaging phantoms with two sequential CT scans, one with and one without a conventional grid. These two scans are processed together to mimic a striped ratio grid. This represents a best case limit of the striped ratio grid, in that the extent of grid ratio modulation is very high and the scatter contrast is maximized. Results: In a uniform cylinder, the striped ratio grid virtually eliminates cupping. Artifacts from scatter are improved in an anthropomorphic phantom. Some banding artifacts are induced by the striped ratio grid. Conclusions: Striped ratio grids could be a simple and effective evolution of conventional antiscatter grids. Construction and validation of a physical prototype remains an important future step.

OSTI ID:
22689393
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 9 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

WE-EF-207-10: Striped Ratio Grids: A New Concept for Scatter Estimation
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22572232

Magnitude and effects of x-ray scatter in a 256-slice CT scanner
Journal Article · Fri Sep 15 00:00:00 EDT 2006 · Medical Physics · OSTI ID:20853456

Improved image quality of cone beam CT scans for radiotherapy image guidance using fiber-interspaced antiscatter grid
Journal Article · Sun Jun 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22412509