Title: Sci-Sat AM: Radiation Dosimetry and Practical Therapy Solutions - 12: Suitability of plan class specific reference fields for estimating dosimeter correction factors for small clinical CyberKnife fields

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4961866· OSTI ID:22689391
; ; ;  [1]
  1. The Ottawa Hospital Cancer Centre, Carleton University, The Ottawa Hospital Cancer Centre, National Research Council (Canada)

Purpose: The purpose of this work is to investigate the utility of plan class specific reference (PCSR) fields for predicting dosimeter response within isocentric and non-isocentric composite clinical fields using the smallest fields employed by the CyberKnife radiosurgery system. Methods: Monte Carlo dosimeter response correction factors (CFs) were calculated for a plastic scintillator and microchamber dosimeter in 21 clinical fields and 9 candidate plan-class PCSR fields which employ the 5, 7.5 and 10 mm diameter collimators. Measurements were performed in 5 PCSR fields to confirm the predicted relative response of detectors in the same field. Results: Ratios of corrected measured dose in the PCSR fields agree to within 1% of unity. Calculated CFs for isocentric fields agree within 1.5% of those for PCSR fields. Large and variable microchamber CFs are required for non-isocentric fields, with differences as high as 5% between different clinical fields in the same plan class and 4% within the same field depending on the point of measurement. Non-isocentric PCSR fields constructed to have relatively homogenous dose over a region larger than the detector have very different ion chamber CFs from clinical fields. The plastic scintillator detector has much more consistent response within each plan class but still require 3–4% corrections in some fields. Conclusions: While the PCSR field concept is useful for small isocentric fields, this approach may not be appropriate for non-isocentric clinical fields which exhibit large and variable ion chamber CFs which differ significantly from CFs for homogenous field PCSRs.

OSTI ID:
22689391
Journal Information:
Medical Physics, Vol. 43, Issue 8; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English