Poster - 44: Development and implementation of a comprehensive end-to-end testing methodology for linac-based frameless SRS QA using a modified commercial stereotactic anthropomorphic phantom

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4961818· OSTI ID:22689350
;  [1]
  1. University of Toronto, Carlo Fidani Peel Regional Cancer Center (Canada)

Purpose: An end-to-end testing methodology was designed to evaluate the overall SRS treatment fidelity, incorporating all steps in the linac-based frameless radiosurgery treatment delivery process. The study details our commissioning experience of the Steev (CIRS, Norfolk, VA) stereotactic anthropomorphic head phantom including modification, test design, and baseline measurements. Methods: Repeated MR and CT scans were performed with interchanging inserts. MR-CT fusion accuracy was evaluated and the insert spatial coincidence was verified on CT. Five non-coplanar arcs delivered a prescription dose to a 15 mm spherical CTV with 2 mm PTV margin. Following setup, CBCT-based shifts were applied as per protocol. Sequential measurements were performed by interchanging inserts without disturbing the setup. Spatial and dosimetric accuracy was assessed by a combination of CBCT hidden target, radiochromic film, and ion chamber measurements. To facilitate film registration, the film insert was modified in-house by etching marks. Results: MR fusion error and insert spatial coincidences were within 0.3 mm. Both CBCT and film measurements showed spatial displacements of 1.0 mm in similar directions. Both coronal and sagittal films reported 2.3 % higher target dose relative to the treatment plan. The corrected ion chamber measurement was similarly greater by 1.0 %. The 3 %/2 mm gamma pass rate was 99% for both films Conclusions: A comprehensive end-to-end testing methodology was implemented for our SRS QA program. The Steev phantom enabled realistic evaluation of the entire treatment process. Overall spatial and dosimetric accuracy of the delivery were 1 mm and 3 % respectively.

OSTI ID:
22689350
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 8 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Technical Note: Evaluation of the systematic accuracy of a frameless, multiple image modality guided, linear accelerator based stereotactic radiosurgery system
Journal Article · Sun May 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22620903

SU-E-T-282: Dose Measurements with An End-To-End Audit Phantom for Stereotactic Radiotherapy
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22548339

Setup Accuracy of the Novalis ExacTrac 6DOF System for Frameless Radiosurgery
Journal Article · Sun Apr 01 00:00:00 EDT 2012 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22056206